Specifications Test specifications The accuracy specification for the display range is defined as ±(%reading + digit counts) at 23 °C ±5 °C, ≤75% RH. Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C. The measurement range meets the service operating errors specified in EN61557-1: 1997, EN61557-2: 1997, EN61557-4: 1997, EN61557-6: 1997, DIN VDE0404-2. Power-on test The test indicates reversed L-N, missing PE, and measures the mains voltage and frequency. Display range 90 V to 264 V Accuracy at 50 Hz ±(2% + 3 counts) Resolution 0.1 V Input impedance > 1 MΩ // 2.2 nF Maximum input mains voltage 264 V Earth bond test (Rpe) Display Range 0 to 19.99 Ω Accuracy (after bond test zeroing) ±(5% + 4 counts) Resolution 0.01 Ω Test current 200 mA AC -0% +40% into 1.99 Ω
10 A AC ±20% into 25 mΩ at 230 VOpen circuit voltage > 4 V AC, < 24 V AC Bond test zeroing can subtract up to 1.99 Ω Insulation test (Riso) Display range 0 to 299 MΩ Accuracy ±(5% + 2 counts) from 0.1 to 300 MΩ Resolution 0.01 MΩ (0 to 19.99 MΩ)
0.1 MΩ (20 to 199.9 MΩ)
1 MΩ (200 to 299 MΩ)Test voltage 500 V DC -0% + 25% at 500 kΩ load or 250 V dc -0% +25% at 250 kΩ load Test current > 1 mA at 500 kΩ load, < 15 mA at 0 Ω Auto discharge time < 0.5 s for 1 μF Max. capacitive load Operational up to 1 μF Touch current test Display range 0 to 1.99 mA AC Accuracy ±(4% + 2 counts) Resolution 0.01 mA Internal resistance (via probe) 2 kΩ Measuring method Probe* Substitute leakage current test Display range 0 to 19.99 mA AC Accuracy ±(2.5% + 3 counts) Resolution 0.01 mA Test Voltage 100 V AC ±20% Load/leakage test: load current Display range 0 A to 13 A Accuracy ±(4% + 2 counts) Resolution 0.1 A* Load/leakage test: load power Display range 230 V Mains 0.0 VA to 3.2 kVA Accuracy ±(5% + 3 counts) Resolution 1 VA (0 to 999 VA), 0.1 kVA (>1.0 kVA)* Load/leakage test: leakage current Display range 0 to 19.99 mA Accuracy ±(4% + 4 counts) Resolution 0.01 mA* PELV test Accuracy at 50 Hz ±(2% + 3 counts) Overload protection 300 V rms Warning threshold 25 V rms RCD test: trip current Operational error ±10% Nominal 30 mA Accuracy ±5% RCD test: trip time Standard requirement 61557 Part 6; tolerance of rated test current 0% to +10% Operational error ±10% RCD type AC general-purpose 30 mA Display range 310 ms Resolution 0.1 ms Accuracy 3 ms Trip time limit at 100% (30 mA) 300 ms Trip time limit at 500% (150 mA) 40 ms *The appliance under test is energized at mains potential. Environmental specifications Operating temperature 0 to 40 °C Relative humidity Non condensing <10 °C 95% from 10 to 30 °C 75% from 30 to 40 °C Safety specifications Safety rating Complies with EN61010-1 3rd edition CAT II, 300 V, pol 2 Mechanical and general specifications Size (L x W x H) 200 x 275 x 114 mm Weight 3.13 kg Power supply 230 V +10% -15%, 50 Hz ±2 Hz or (6500-2 only: 110V +10% -15%,50 Hz ±2 Hz) Power consumption (tester) 13 W typical (idle)
60 W max. during 25A Bond testStorage Temperature -10 to 60 °C Corrosion 70 °C @ 95% RH for 5 days max. Operating altitude 0 up to 2000 m Sealing IP-40 (enclosure), IP-20 (connectors) EMC Complies with EN61326-1, portable EMI Immunity 3 V/m