  • Differential Module Specifications
    Model Parameter / range Burst rating6 Hi side media2 Lo side media2
    750P00 0 to 1 inH2O (0 to 2.5 mBar) 30X Dry air Dry air
    750P3IN 0 to 3 Inch H2O (0 to 7.5 mbar) 10X Dry air Dry air
    750P5IN 0 to 5 Inch H2O (0 to 12.5 mbar) 6X Dry air Dry air
    750P017 0 to 10 inH2O (0 to 25 mBar) 3X Dry air Dry air
    750P02 0 to 1 psi (0 to 70 mBar) 3X Dry air Dry air
    750P22 0 to 1 psi (0 to 70 mBar) 3X 316 SS Dry air
    750P03 0 to 5 psi (0 to 350 mBar) 3X Dry air Dry air
    750P23 0 to 5 psi (0 to 350 mBar) 4X 316 SS Dry air
    750P04 0 to 15 psi (0 to 1 bar) 3X Dry air Dry air
    750P247 0 to 15 psi (0 to 1 bar) 4X 316 SS Dry air
    Model Reference uncertainty4 Total uncertainty
    1-year
    (15-35°C)
    Total uncertainty
    1-year1
    Total uncertainty
    6-month
    (15-35°C)
    Total uncertainty
    6-month1
    750P00 ±0.15% ±0.3% ±0.35% ±0.25% ±0.30%
    750P3IN ±0.15% ±0.3% ±0.35% ±0.25% ±0.30%
    750P5IN ±0.15% ±0.3% ±0.35% ±0.25% ±0.30%
    750P017 ±0.1% ±0.2% ±0.3% ±0.15% ±0.25%
    750P02 ±0.050% ±0.1% ±0.15% ±0.075% ±0.125%
    750P22 ±0.050% ±0.1% ±0.15% ±0.075% ±0.125%
    750P03 ±0.02% ±0.04% ±0.05% ±0.035% ±0.04%
    750P23 ±0.02% ±0.04% ±0.05% ±0.035% ±0.04%
    750P04 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P247 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    Gage Module Specifications
    Model Parameter / range Burst rating6 Hi side media2 Lo side media2
    750P057 0 to 30 psi (0 to 2 bar) 4X 316 SS N/A
    750P067 0 to 100 psi (0 to 7 bar) 4X 316 SS N/A
    750P277 0 to 300 psi (0 to 20 bar) 4X 316 SS N/A
    750P07 0 to 500 psi (0 to 35 bar) 4X 316 SS N/A
    750P08 0 to 1000 psi (0 to 70 bar) 3X 316 SS N/A
    750P097 0 to 1500 psi (0 to 100 bar) 3X 316 SS N/A
    750P2000 0 to 2000 psi (0 to 140 bar) 3X 316 SS N/A
    Model Reference uncertainty4 Total uncertainty
    1-year
    (15-35°C)
    Total uncertainty
    1-year1
    Total uncertainty
    6-month
    (15-35°C)
    Total uncertainty
    6-month1
    750P057 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P067 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P277 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P07 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P08 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P097 ±0.0175% ±0.035% ± 0.045% ±0.03% ±0.04%
    750P2000 ±0.0175% ±0.035% ± 0.045% ±0.03% ±0.04%
    High Pressure Module Specifications
    Model Parameter / range Burst rating6 Hi side media2 Lo side media2
    750P29 0 to 3000 psi (0 to 200 bar) 3X 316 SS N/A
    750P30 0 to 5000 psi (0 to 340 bar) 3X 316 SS N/A
    750P31 0 to 10000 psi (0 to 700 bar) 2X 316 SS N/A
    Model Reference uncertainty4 Total uncertainty
    1-year
    (15-35°C)
    Total uncertainty
    1-year1
    Total uncertainty
    6-month
    (15-35°C)
    Total uncertainty
    6-month1
    750P29 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P30 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    750P31 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
    Absolute Module Specifications
    Model Parameter / range Burst rating6 Hi side media2 Lo side media2
    750PA3 0 to 5 psia (0 to 350 mBar) 4X 316 SS N/A
    750PA47 0 to 15 psia (0 to 1 bar) 4X 316 SS N/A
    750PA5 0 to 30 psia (0 to 2 bar) 4X 316 SS N/A
    750PA6 0 to 100 psia (0 to 7 bar) 4X 316 SS N/A
    750PA27 0 to 300 psia (0 to 20 bar) 4X 316 SS N/A
    750PA7 0 to 500 psia (0 to 35 bar) 4X 316 SS N/A
    750PA8 0 to 1000 psia (0 to 70 bar) 3X 316 SS N/A
    750PA9 0 to 1500 psia (0 to 100 bar) 3X 316 SS N/A
    Model Reference uncertainty4 Total uncertainty
    1-year
    (15-35°C)
    Total uncertainty
    1-year1
    Total uncertainty
    6-month
    (15-35°C)
    Total uncertainty
    6-month1
    750PA3 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PA47 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PA5 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PA6 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PA27 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PA7 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PA8 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PA9 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    Vacuum Module Specifications
    Model Parameter / range Burst rating6 Hi side media2 Lo side media2
    750PV3 -5 psi (-350 mBar) 4X 316 SS Dry air
    750PV4 -15 psi (-1 bar) 4X 316 SS Dry air
    Model Reference uncertainty4 Total uncertainty
    1-year
    (15-35°C)
    Total uncertainty
    1-year1
    Total uncertainty
    6-month
    (15-35°C)
    Total uncertainty
    6-month1
    750PV3 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PV4 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    Dual Range Specifications
    Model Parameter / range Burst rating6 Hi side media2 Lo side media2
    750PD2 -1 to 1 psi (-70 to 70 mBar) 4X 316 SS Dry air
    750PD3 -5 to 5 psi (-350 to 350 mBar) 4X 316 SS Dry air
    750PD10 -10 to 10 psi (-700 to 700 mBar) 4X 316 SS Dry air
    750PD4 -15 to 15 psi (-1 to 1 bar) 4X 316 SS Dry air
    750PD5 -15 to 30 psi (-1 to 2 bar) 4X 316 SS
    750PD50 -15 to 50 psi (-1 to 3.5 bar) 4X 316 SS
    750PD6 -15 to 100 psi (-1 to 7 bar) 4X 316 SS
    750PD7 -15 to 200 psi (-1 to 14 bar) 4X 316 SS
    750PD27 -15 to 300 psi (-1 to 20 bar) 4X 316 SS
    Model Reference uncertainty4 Total uncertainty
    1-year
    (15-35°C)
    Total uncertainty
    1-year1
    Total uncertainty
    6-month
    (15-35°C)
    Total uncertainty
    6-month1
    750PD2 ±0.05% ±0.1% ±0.15% ±0.075% ±0.125%
    750PD3 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
    750PD10 ±0.025% ±0.05% ±0.07% ±0.04% ±0.06%
    750PD4 ±0.0175% ±0.035% ±0.045% ±0.03% ± 0.04%
    750PD5 ± 0.0175% ± 0.035% ± 0.045% ±0.03% ± 0.04%
    750PD50 ± 0.0175% ± 0.035% ± 0.045% ±0.03% ± 0.04%
    750PD6 ± 0.0175% ± 0.035% ± 0.045% ±0.03% ± 0.04%
    750PD7 ± 0.0175% ± 0.035% ± 0.045% ±0.03% ± 0.04%
    750PD27 ± 0.0175% ± 0.035% ± 0.045% ±0.03% ± 0.04%
    Reference Module Specifications
    Model Parameter / range Burst rating6 Hi side media2 Lo side media2
    750R045 0 to 15 psi (0 to 1 bar) 3X Dry air Dry air
    750R065 0 to 100 psi (0 to 7 bar) 4X 316 SS
    750R27 0 to 300 psi (0 to 20 bar) 4X 316 SS
    750R07 0 to 500 psi (0 to 35 bar) 4X 316 SS
    750R085 0 to 1000 psi (0 to 70 bar) 3X 316 SS
    750R29 0 to 3000 psi (0 to 200 bar) 3X 316 SS
    750R30 0 to 5000 psi (0 to 340 bar) 3X 316 SS
    750R315 0 to 10000 psi (0 to 700 bar) 2X 316 SS
    750RD5 -15 to 30 psi (-1 to 2 bar) 4X Dry air
    750RD65 -12 to 100 psi (-0.8 to 7 bar) 4X 316 SS
    750RD27 -12 to 300 psi (-0.8 to 20 bar) 4X 316 SS
    Model Reference uncertainty4 Total uncertainty
    1-year
    (15-35°C)
    Total uncertainty
    1-year1
    Total uncertainty
    6-month
    (15-35°C)
    Total uncertainty
    6-month1
    750R045 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750R065 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750R27 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750R07 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750R085 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750R29 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750R30 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750R315 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750RD5 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750RD65 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    750RD27 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
    1. Total uncertainty, % of full span for temperature range 0°C to +50°C, one year interval. Total uncertainty, 1.0% of full span for temperature range -10°C to 0°C, one year interval. No 6 month specification available for range -10°C to 0°C.
    2. “Non-corrosive gasses” indicates Dry air or non-corrosive gas as compatible media. “Stainless Steel 316-SS” indicates media compatible with Type 316 Stainless Steel.
    3. Specifications % of Full Span unless otherwise noted.
    4. Reference Uncertainty is the specification for as left data for 24 hours.
    5. When reference class modules are used with fixed resolution products (717, 718, 719 series, 725 and 726) calibrators add ±1 count to the overall accuracy specification.
    6. Burst rating specification refers to the multiplier times full scale of the module for the rated burst pressure.
    7. Intrinsically Safe pressure module available in this pressure range.

Videó

Specifikáció

Letölthető termék információ:  Fluke 750P

Ajánlatkérés

Célunk, hogy minél egyszerűbbé tegyük  az Ön számára az ajánlatkérést.

Kérjük, töltse ki az alábbi űrlapot és küldje el.

Kollégáink hamarosan jelentkezni fognak az ajánlattal.

Ajánlatkérés

Telefonos visszahívás kérése

Célunk, hogy minél egyszerűbbé tegyük  az Ön számára a termékeinkkel kapcsolatos információhoz jutást.

Kérjük, töltse ki az alábbi űrlapot és küldje el.

Kollégáink hamarosan jelentkezni fognak, hogy megválaszolják kérdését.

Telefonos visszahívás