Leírás
AC WITHSTANDING |
|||
Output-Voltage Range | 0.050kV~ 5.000kV ac | ||
Output-Voltage Resolution | 2V/step | ||
Output-Voltage Accuracy | ± (1% of setting +5V) [no load] | ||
Maximum Rated Load | 500 VA (5kV/100mA) | ||
Maximum Rated Current | 100mA (0.5kV < V ≦ 5kV)
10mA (0.05kV ≦ V≦ 0.5kV) |
||
Output-Voltage Waveform | Sine wave | ||
Output-Voltage Frequency | 50 Hz / 60 Hz selectable | ||
Voltage regulation | ± (1% of reading +5V) [full load → no load] | ||
Voltmeter Accuracy | ± (1% of reading+ 5V) | ||
Current Measurement Range | 0.001mA~100.0mA | ||
Current Best Resolution | 0.001mA / 0.01mA / 0.1mA | ||
AC Current Measurement Accuracy | ±(1.5% of reading +30 counts) when HISET < 1.11mA
±(1.5% of reading +3 counts) when HISET ≧ 1.11mA |
||
Window Comparator Method | Yes | ||
ARC Detect | Yes | ||
RAMP (Ramp-Up Time) | 0.1~999.9s | ||
TIMER (Test Time)* | OFF, 0.5s~999.9s | ||
Sweep Function* | Yes | ||
GND | ON/OFF | ||
DC WITHSTANDING |
|||
Output-Voltage Range | 0.050kV~ 6.000kV dc | ||
Output-Voltage Resolution | 2V/step | ||
Output-Voltage Accuracy | ± (1% of setting +5V) [no load] | ||
Maximum Rated Load | 100W (5kV/20mA) | ||
Maximum Rated Current | 20mA (0.5kV < V ≦ 6kV)
2mA (0.05kV ≦ V≦ 0.5kV) |
||
Voltage regulation | ± (1% of reading +5V) [full load → no load] | ||
Voltmeter Accuracy | ± (1% of reading+ 5V) | ||
Current Measurement Range | 0.001mA~020.0mA | ||
Current Best Resolution | 0.001mA / 0.01mA / 0.1mA | ||
DC Current Measurement Accuracy | ±(1.5% of reading +30 counts) when HISET < 1.11mA
±(1.5% of reading +3 counts) when HISET ≧ 1.11mA |
||
Window Comparator Method | Yes | ||
ARC Detect | Yes | ||
RAMP (Ramp-Up Time) | 0.1~999.9s | ||
TIMER (Test Time)* | OFF, 0.5s~999.9s | ||
Sweep Function* | Yes | ||
GND | ON/OFF | ||
INSULATION RESISITANCE |
|||
Output-Voltage | 50V~1000V dc | ||
Output-Voltage Resolution | 50V/step | ||
Output-Voltage Accuracy | ± (1% of setting +5V) [no load] | ||
Resistance Measurement Range | 0.001 Gohm~ 50.00 Gohm | ||
Test voltage | Measurable Range | Accuracy | |
50V≦V≦450V | 0.001~ 0.050 Gohm
0.051~ 2.000 Gohm |
±(5% of reading +1 count)
±(10% of reading +1 counts) |
|
500V≦V≦1000V | 0.001~ 0.500 Gohm
0.501~ 9.999 Gohm 10.00~ 50.00 Gohm |
±(5% of reading +1 count)
±(10% of reading +1 count) ±(20% of reading +1 count) |
|
Window Comparator Method | Yes | ||
Output Impedance | 600 kohm | ||
RAMP (Ramp-Up Time) | 0.1~999.9s | ||
TIMER (Test Time) | 1s~999.9s | ||
Sweep Function* | Yes | ||
GND | OFF (fixed) | ||
GROUND BOND |
|||
Output-Current Range | 03.00A~32.00A ac | ||
Output-Current Resolution | 0.01A | ||
Output-Current Accuracy | 3A≦I≦8A : ± (1% of setting +0.2A)
8A<I≦32A : ± (1% of setting +0.05A) |
||
Test-Voltage | 6Vac max. (open circuit) | ||
Test-Voltage Frequency | 50 Hz / 60 Hz selectable | ||
Resistance Measurement Range | 10mW~650.0m ohm | ||
Resistance Measurement Resolution | 0.1mW | ||
Resistance Measurement Accuracy | ± (1% of reading +2m ohm) | ||
Window Comparator Method | Yes | ||
TIMER (Test Time) | 0.5s~999.9s | ||
Sweep Function* | Yes | ||
GND | OFF (fixed) | ||
Test Method | Four Terminal | ||
MEMORY |
|||
Single Step Memory | MANU : 100 blocks | ||
Automatic Testing Memory | AUTO : 100 blocks, menu per auto : 16 | ||
INTERFACE |
|||
Rear Output | Standard | ||
RS232 | Standard | ||
USB (Device) | Standard | ||
GPIB | Option | ||
Remote Terminal (Front) | Standard | ||
SIGNAL IO | Standard | ||
General |
|||
DISPLY | 240 x 64 dot matrix LED back light LCD | ||
POWER SOURCE | AC100V/120V/220V/230V ±10%, 50/60Hz | ||
POWER CONSUMPTION | 1000VA max. | ||
DIMENSIONS & WEIGHT | 330(W) x 148(H) x 587(D) mm (GPT-9904/9903),
330(W) x 148(H) x 482(D) mm (GPT-9903A/9902A/9901A) approx.27kg(Max) |