Fluke 707Ex Robbanásbiztos hurok kalibrátor

Description

Specifications
Current measurement Range 0 to 24 mA
Resolution 0.001 mA
Accuracy 0.015% Rdg + 2 LSD
Current sourcing Range 0 to 20 mA or 4 to 20 mA
Accuracy 0.015% Rdg + 2 LSD
Drive capability 700 Ω at 20 mA
Current simulation Range 0 to 20 mA or 4 to 20 mA
Accuracy 0.015% Rdg + 2 LSD
Loop power While measuring mA, 24 V
Voltage measurement Range 0 to 28 V DC
Resolution 1 mV
Accuracy 0.015% Rdg + 2 LSD
Environmental Specifications
Operating temperature -10ºC to 55ºC
Storage temperature -40ºC to 60ºC
Humidity (without condensation) 0 to 95% (0ºC to 30ºC);
0 to 75% (30ºC to 40ºC)
0 to 45% (40ºC to 50ºC)
Temperature coefficient -10ºC to 18ºC, 28ºC to 55ºC, ±.005% of range per ºC
Safety Specifications
Safety CSA C22.2 No. 1010.1:1992
EMC EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage 30 V
ATEX certification Ex II G Ex ia IIC T4 Zones 1 and 2, ZELM 02 ATEX 0120 X
CE certification EN61010-1 and EN61326
Mechanical and General Specifications
Size 152 x 70 x 41 mm
Weight 224 g (350 g including holster)
Batteries One 9V battery Alkaline
Warranty 3 year
Shock & vibration per MIL-T-28800 for a Class 2 Instrument
Display current and % of span Yes
Auto step, auto ramp Yes
Span check Yes